Remote Controlled Fluoroscopic X-ray Unit GX100 |
| Application: |
This unit is suitable for fluoroscopy on human body, esophagus, lung, chest, abdomen, limbs and pelvis cavity. It can
perform routine radiography also. |
| |
| Composition: |
- Control console
- H.T generator: inversion type
- Diagnostic body.
- Image intensifier and monitor
|
| |
Features |
- High frequency and intelligent controlled generator, LED display, error code display and function of both fluoroscopy and radiography.
- Remote motor driven diagnostic body, large movement range on direction of up and down, transverse and rotation, easy operation.
- TV system with high definition and strong function of digital image processing to acquire excellent images.
- Auto. Brightness Control, suitable for team physical examination.
- Humanity design, meet to doctor’s different requirement
|
| |
| Specification: |
·Diagnostic Body:
- Range of up and down movement: 870 ~ 1750mm
- Range of transverse movement: ±125mm
- Range of rotating movement: ±45°
- SID: 1m
- Collimator: electric driven
- Cassette size: 10” x 12”
·H.T generator
| |
1. Max. out put power: 4kW |
| |
2. Main inversion frequency: ≥25kHz |
| |
3. Tube voltage: 40kV ~110kV |
| |
4. Tube current: |
| |
fluoroscopy: 0.3 ~4mA
radiography: 10 ~63mA |
| |
5. Radiographic time: 3.2ms ~3.2s |
| |
6. X-ray tube: small focus: 1.0mm, large focus: 2.6mm |
| |
|
·TV system
- Image intensifier: ¢230mm (9")
- Camera: 1/2” CCD, 570TVL, CCIR
- Resolution: more than 12LP/cm,gray level: ≥8
- Image processing: LIH, 8 image storage, noise reduction can be continue adjustment. AGC, Gray level revise. Image reverse. Negative image.
- Monitor: 14”, 1000TVL
·Power supply:
- Voltage: single phase AC 220V±22
- Frequency: 50Hz±1Hz
- Inner resistance: ≤0.5Ω
- Capacity: ≥6kVA
|
| |
|
|