SC-300 versatile ventilator serves as long term ventilator for respiratory rescue and therapy,which is pneumatic driven,electronic controlled suitable for patients of all ages (from prematures to adults).SC-300 ventilator has the superior performance and functions as a modern ventilator. SC-300 ventilator adopts a dual-microprocessor system and integrates most of modern ventilation modes including hundred percent oxygen enrichment,sigh function and medicament nebulization.It also has monitoring functions with audible and visual alarm.To make it easy for emergency use the SC-300 ventilator has an additional key for "emergency".When use the emergency function,all that required is to set a O2 concentration and aspiratory pressure. With a well designed structure,new ventilation mode can be easily added to SC-300 ventilator by upgrading its software without changing any hardware. |
• For patient of all ages
• Pneumatic driven, electronic controlled.Wide selection of ventilation modes.Volume constant, Pressure control,Pressure
support etc.
• Dual-microprocessor system,high reliability.
• Monitoring of airway pressure and ventilation volume.
• Advertisement of operation for easy use.
• Build-in alarm function for safety. |
• Ventilation frequency:
IPPV:5~60bpm SIMV:1~12bpm
• Inspiratory flow:
Infant:4~ 24L /min(continuous)
Adult:10~ 120L /min(inspiratory phase)
• Inspiratory time:0.2~3s
• Insperatory pause:
0 -2s[for IPPV(VC) and SIMV]
• Peak pressure:1~9kPa
• Inspiratory oxygen concentration:21~100%
• PEEP/CPAP:0~1.5kPa
• Pressure support:0~3.5kPa
• Trigger sensitivity: -0.1~1.0kPa
• Ventilation modes: Adult:IPPV(VC),IPPV(PC),SIMV,SIMV+PSV,CPAP,PSV Infant:IPPV(PC),IMV,CPAP,Pressure Limit,Time Cycle. Emergency:IPPV(PC)
(02 concentration, inspiratory pressure key active only)
• MV monitoring: Upper Limit:5 - 40L /min Lower Limit:0~ 20L /min
• Airway pressure: monitoring:Pmax,Pmean,Pplat,PEEP
• Ventilation monitoring: Frequency,Minute Volume,Tidal Volume
• Alarms:Air supply deficiency,Oxygen supply deficiency,Hose detachment,Airway pressure high,Airway pressure low,MV
high,MV low,Apnea,Power failure
|