MHJ-IIIB2 multifunction anaesthesia machine, developed with reliability, safety, practicability, combination in mind,is a kind of up to date economic anaesthesia machine with modular system.Standard parts of MHJ-IIIB2 include anaesthesia ventilator SC-M3A,a compact vaporizer and airway gauge.It is possible to be furnished with other types of vaporizers and electronic monitorings to make an ideal multifunction anaesthesia machines for modern hospital. |
• ISOcircuit system adaptable to various monitoring
• New type of flowmeter ORC system (O2concentration no less than 25%),N2O cut off safety unit.
• Conpact Enflurane Vaporizer:Vaporized drug comcentration adjustable
• The SC-M3A pneumatic anaesthesia ventilator was comparable small consumption of gas(one cylinder of Oxygen for six hours of operation),and precise parameters for control
• Airway pressure gauge for real-time display of respiration of patient
|
• Flow control range:O2 (0.1~10L/min)
N2O(0.1~10L/min)
• Gas Supply:0.27~0.55MPa
• O2Flush:35~70L/min
• N2O automatic cut-off:active when O2 pressure lower than 0.1MPa
• Non-compensation Vaporizer(Enflurane), Cncentration adjustable(0~5%)
• Anaesthesia Ventilator:
Control type:Pneumatically driven and
Controlled,time cycle and volume constant
Gas supply:0.27~0.55Mpa
Frequency:6~40cycles/min
I:E ratio:1:2(Fixes)
Tidal volume: Adule bellow 150~1600ml
Children bellow 50~150ml
Safety Valve Working Pressue:6kPa
|