It is suitable for both fluoroscopy and radiography in hospitals and health centers. It can also be used in conventional stomach examinations. The unit is provided with a spot-film device, a motor-driven Bucky, a collimator and a motor-driven diagnostic table with the tabletop movable. It is an ideal x-ray diagnostic unit with multi-function
MAIN TECHNICAL PARAMETER |
X-ray generator |
Rotary anode, single focus, and bridge silicon rectification, The X-ray generator is an oil-immersed self-cooled unit. |
Max rated capability |
Fluoroscopy |
75kVp . 3mA |
Radiography |
Current |
15 mA |
30 mA |
60 mA |
100 mA |
Voltage |
90 kVp |
90 kVp |
90 kVp |
80 kVp |
Time |
6.3s |
6.3s |
2.0s |
1.0s |
Power supply |
Voltage |
180 ~ 240V |
Frequency |
50Hz |
Power |
not lower than 7kVA |
Resistance |
not larger than 1Ω |
Voltage adjustment range of x-ray tube |
45 ~ 90kVp (continuous) |
Range of the timer |
0.08 ~ 6.3 s |
Fluorescent screen size |
356 mm × 356 mm |
Distance between the focus to the screen |
830 mm (max)
630 mm (min) |
Tube head carriage movement range |
vertical: 1100 mm
horizontal: 260 mm |
Specification of X-ray tube |
XD55 0.8mm × 0.8mm
Rotary anode, single focus |
Collimator. |
Field coverage at 100 cm FFD . 45× 45 cm max,0× 0 cm min |
Movement of the tabletop |
400 mm |
Field size of spot film device |
127mm × 178mm (one piece)
203mm × 254mm (one piece)
1/2 ( 127mm × 178 mm ) (two pieces) |
Transport dimension . Weight |
234×100×123(cm) . NET: 410kg /GROSS: 535kg |